Note:
After retrieving the settings in Levelogger Software, If the Wellhead batteries have drained to 20% charge or lower, a warning message will appear to replace the batteries.
Each Wellhead uses four (4) 1.5V AA replaceable lithium batteries. To install the batteries, follow the steps below:
- Unscrew the top of the Wellhead to access the battery compartment.
- Carefully remove the battery holder from the compartment.
- Install the four batteries in the holder. Ensure proper polarity when placing the batteries, paying attention to the positive {+} and negative {-} symbols etched in the holder.
- Slide the battery holder back into the compartment. Ensure proper alignment; the spring contacts in the housing should be lined up with the protruding contacts on the bottom of the battery holder.
- Screw the top cap firmly back onto the Wellhead.
- After replacing the batteries, to reset the battery indicator in the Levelogger Software, use the Diagnostic Utility (see Section 10.6).
Figure 8-3 Installing Wellhead Batteries
8.2.1.1 Wellhead Battery Life Estimates
AquaVent Sampling Rate | Battery Life Estimate |
---|---|
8 readings per second | 10 days |
1 reading per second | 3 months |
1 reading per minute | 8 years |