6.7 Vented Dataloggers Zero

In the Dataloggers Screen, there is a Zero Barometric option to zero a LevelVent 5 or AquaVent 5 vented datalogger to current atmospheric pressure. This will ensure that it is reading 0 in air for the level reading.

setting zero barometric pressure for solinst vented dataloggers on ios levelogger app

Figure 6-28 Zero Barometric – iOS

setting zero barometric pressure for solinst vented dataloggers on android levelogger app

Figure 6-29 Zero Barometric – Android


On an iOS device, after selecting Zero Barometric and clicking "OK", the current date and time will be displayed as the Datalogger Zero Date in the Dataloggers Settings screen. You can also select Zero Barometric in the Dataloggers Settings screen to zero the datalogger.

On an Android device, after selecting Zero Barometric a message will appear stating "Zero Offset Calibrated" and the current date and time will be displayed as the Zero Offset in the Diagnostic Utility (See Section 11).

 

setting zero barometric data on solinst levelogger app for ios

Figure 6-30 Datalogger Zero Date – iOS

setting zero barometric data on solinst levelogger app for android

Figure 31 Zero Offset Date – Android

 

Note:

You can not perform a Datalogger Zero while the datalogger is running.