The Barologger Gold uses algorithms based on air pressure only. It measures and logs changes in atmospheric pressure, which are then used to compensate water level readings recorded by a Levelogger.
Note:
Solinst recommends using the most recent Barologger Gold firmware version with Levelogger Software Version 4.4.0. See Section 2.3.
Barologger Gold Technical Specifications | |
---|---|
Level Sensor: | Piezoresistive Silicon in 316L Stainless Steel |
Accuracy: | 0.05% FS |
Normalization: | Automatic Temperature Compensation |
Temp. Comp. Range: | 0ºC to 40°C |
Temperature Sensor: | Platinum Resistance Temperature Detector (RTD) |
Temp. Sensor Accuracy: | ± 0.05°C |
Temp. Sensor Resolution: | 0.003°C |
Battery Life: | 10 Years - based on one reading/min |
Clock Accuracy (typical): | ± 1 minute/year |
Operating Temperature: | -20°C to 80°C |
Maximum # Readings: | 40,000 of pressure and temperature |
Memory: | Slate and Continuous |
Communication: | Optical Infra-red: USB, RS-232, SDI-12 |
Communication Speed: | 9600 bps |
Size: | 22 mm x 154 mm (7/8" x 6" ) |
Weight: | 179 grams (6.3 oz) |
Corrosion Resistance: | Zirconium Nitride (ZrN) Coating |
Other Wetted Materials: | 316L Stainless Steel, Delrin®, Viton® |
Sampling Modes: | Linear, Event & User-Selectable with 30 separate line items, Future Start, Future Stop, Real Time View |
Measurement Rates: | 0.5 sec to 99 hrs |
LT Models | Full Scale (FS) | Accuracy |
---|---|---|
Barologger | Air Only | ± 0.003 ft., 0.1 cm |