- Make sure other SDI-12 devices are not being addressed and communicating at the same time on the SDI-12 network.
- Using your SDI-12 datalogger and its SDI-12 software, issue an Acknowledge Active Command (a!), to test if the device address is correct and you can communicate with the newly installed SDI-12 Interface Cable and Levelogger.
- Issue the Start Verification Command (aV!) and later read it with the Send Data Command (aD0!) to verify the Levelogger and the SDI-12 Interface Cable are operating without errors.
- Use the Send Identification Command (aI!) to verify the name and model of the Levelogger matches that of the Levelogger you have just installed. Execute this command 2-3 times to ensure the Levelogger identification information gets read and registered with the SDI-12 Interface Cable.
- Execute the Start Concurrent Measurement Command (aC!) and later retrieve the measurement with the Send Data Command (aD0!), to ensure the Levelogger is connected and properly taking measurements.
- If the above steps are successful, you may program the datalogger to automatically issue commands to the SDI-12 Interface Cable and Levelogger, to take measurements and record them.