Lightweight Water Level Meter Operating Instructions
Spohr-Messtechnik GmbH
Länderweg 37
D-60599 Frankfurt a.M.
Germany
Tel: 069 622860
Fax: 069 620455
[email protected]
Spohr Lightweight Water Level Meter
Front View
- Handle
- Indicator Light
- Buzzer
- Battery Drawer
Spohr Lightweight Water Level Meter
Back View
- Brake
Equipment Check
Submerse the probe in tap water. This completes the circuit and activates the buzzer and light.
Water Level Measurements
- The probe zero measurement point is the tip of the needle in the center of the probe.
- Feed the tape into and out of the well. The light and buzzer activate when the zero point enters water. To ensure accuracy, lower and raise the probe a few times and then record the depth measurement from the tape at the top of the well.
Routine Care
- After the depth to water has been recorded, the tape should be carefully rewound onto the reel, the probe wiped dry.
- The probe, tape and reel can be cleaned with phosphate free (non-abrasive) detergent and warm water. Do not submerge the reel.
- Use of a carrying bag adds to the service life of the meter.
Probe Care and Cleaning
- The probe should be wiped clean after each use.
- Remove any dirt or water from around the central sensor pin.
- Check the probe seal/strain relief for any cracking or other damage.
Battery Replacement
Battery type – alkaline, 9 volt.
- The battery is housed in a convenient battery drawer located in the faceplate of the Water Level Meter.
- To replace the battery, simply press the drawer in, lift then pull.
- The battery drawer should slide out of the faceplate enough to pull it out.
- Note the polarity (positive (+) terminal should be towards the small notch in the end of the drawer) and place new battery in the drawer and slide it back into the faceplate.
Symptom
Cause
Remedy
No sound when probe immersed in water.No sound when probe immersed in water.
Dead battery.
Replace with 9V Alkaline.
Symptom
Cause
Remedy
Disconnected wires on circuit board.
Check all connections inside hub of reel for loose/disconnected wires - solder or reconnect.
Symptom
Cause
Remedy
Broken wire in tape.
Locate break in tape - splice and seal, or replace. (Contact Spohr)
Symptom
Cause
Remedy
Disconnected wire inside probe.
Contact Spohr for instructions.
Symptom
Cause
Remedy
Instrument continuously sounds after being immersed in water.
Water in probe. Probe may be dirty which could interfere with the circuit connection.
Clean probe. Contact Spohr for instructions.